LATIN AMERICAN IMAGE QUALITY SURVEY IN DIGITAL MAMMOGRAPHY STUDIES

Nenhuma Miniatura disponível
Data
2017
Autores
Mora, Patricia
Khoury, Helen
Bitelli, Regina [UNIFESP]
Rosa Quintero, Ana
Garay, Fernando
Garcia Aguilar, Juan
Gamarra, Mirtha
Ubeda, Carlos
Orientadores
Tipo
Artigo
Título da Revista
ISSN da Revista
Título de Volume
Resumo
Under International Atomic Energy Agency regional programme TSA3 Radiological Protection of Patients in Medical Exposures, Latin American countries evaluated the image quality and glandular doses for digital mammography equipment with the purpose of seeing the performance and compliance with international recommendations. Totally, 24 institutions participated from Brazil, Chile, Costa Rica, El Salvador, Mexico, Paraguay and Venezuela. Signal difference noise ratio results showed for CR poor compliance with tolerances
better results were obtained for full-field digital mammography equipment. Mean glandular dose results showed that the majority of units have values below the acceptable dose levels. This joint Latin American project identified common problems: difficulty in working with digital images and lack of specific training by medical physicists from the region. Image quality is a main issue not being satisfied in accordance with international recommendations
optimisation processes in which the doses are increased should be very carefully done in order to improve early detection of any cancer signs.
Descrição
Citação
Radiation Protection Dosimetry. Oxford, v. 174, n. 1, p. 94-101, 2017.