Sistema eletrônico para validação da aferição e calibração dos tonômetros tipo Goldmann

View/ Open
Date
2016-09-27Author
Oliveira, Amaro Aparecido de [UNIFESP]
Advisor
Hirai, Flavio Eduardo [UNIFESP]Type
Dissertação de mestrado profissionalMetadata
Show full item recordAbstract
The purpose of this study was to develop a digital device for the assessment of accuracy of the Goldmann tonometer for calibration purposes. The current method uses a graded metal rod attached to the tonometer where a correlation between its displacement and the pressure exerted by the human eye is verified. The physical variable (pressure) was measured by an indirect non-invasive sensor. Analog signals received by the equipment were converted into digital information and shown in an alphanumeric display. This study showed the potential development of a digital device for calibration of Goldmann tonometers. The results may also serve as basis for creation of new policies for measurement control and calibration of tonometers in Brazil. O objetivo deste estudo foi desenvolver dispositivo digital para verificação da acurácia do método atual de aferição de tonômetros de aplanação de Goldmann. Atualmente o método utilizado é a checagem através de uma haste metálica graduada conectada ao tonômetro, onde é estabelecida uma correlação entre o deslocamento da mesma e a pressão exercida pelo olho do indivíduo no instrumento de medição. A grandeza foi medida através de um elemento sensor indireto, não invasivo e sem contato com o paciente. Os sinais captados analógicos foram posteriormente convertidos em informação digital visível em mostrador tipo display alfanumérico. O estudo demonstrou a possibilidade de desevolvimento de um dispositivo digital para utilização na calibração de tonômetros do tipo Goldmann. Os resultados obtidos também podem servir como base para criação de normas específicas para controle metrológico deste equipamento no Brasil.