Repeatability of central corneal thickness measurement with the Pentacam HR system

View/ Open
Date
2012-01-01Author
Alonso, Ruiz Simonato
Fontes, Bruno Machado [UNIFESP]
Ventura, Marcelo Palis
Ambrósio, Renato [UNIFESP]
Type
ArtigoISSN
0034-7280Is part of
Revista Brasileira De OftalmologiaDOI
10.1590/S0034-72802012000100003Metadata
Show full item recordAlternative Title
Reprodutibilidade da medida da espessura central da córnea com o sistema Pentacam HRAbstract
Purpose: To assess the repeatability of central corneal thickness measurement at the geometrical center (Central Corneal Thickness - CCT) given by the Pentacam High Resolution (HR) Comprehensive Eye Scanner (Oculus, Wetzlar, Germany) over time. Methods: Prospective, single center, observational study. Two separate CCT measurements were taken by the Pentacam corneal tomography exam (CTm) 3 to 12 months apart, and compared. Results: One hundred and sixteen eyes (n=116) of 62 health patients were included in this study. Average CCT in first and last visits was 541.6 +/- 37 mu m and 543.6 +/- 36.9 mu m respectively. Mean difference between both measurements was 9.2 +/- 6.4 mu m, and there was no statistically significant difference in CCT measurement between visits, with good correlation between them (P = 0.057, r(2) = 0,9209). Conclusion: Pentacam (HR) CTm gives repeatable CCT measurements over time.
Citation
Revista Brasileira De Oftalmologia. Rio De Janeiro: Soc Brasileira Oftalmologia, v. 71, n. 1, p. 14-17, 2012.Keywords
Cornea/anatomy & histologyCorneal topography
Biometry/methods
Reproducibility of results
Diagnostic techniques, ophthalmological/instrumentation