Title: Automated Quantitation of Choroidal Neovascularization: A Comparison Study Between Spectral-Domain and Swept-Source OCT Angiograms
Authors: Zhang, Qinqin
Chen, Chieh-Li
Chu, Zhongdi
Zheng, Fang
Miller, Andrew
Roisman, Luiz [UNIFESP]
Dias, Joao Rafael de Oliveira
Yehoshua, Zohar
Schaal, Karen B.
Feuer, William
Gregori, Giovanni
Kubach, Sophie
An, Lin
Stetson, Paul F.
Durbin, Mary K.
Rosenfeld, Philip J.
Wang, Ruikang K.
Keywords: optical coherence tomography
swept-source OCTA
spectral-domain OCTA
neovascular AMD
macular neovascularization
pigment epithelium detachment
Issue Date: 2017
Publisher: Assoc Research Vision Ophthalmology Inc
Citation: Investigative Ophthalmology & Visual Science. Rockville, v. 58, n. 3, p. 1506-1513, 2017.
Abstract: PURPOSE. To compare the lesion sizes of choroidal neovascularization (CNV) imaged with spectral-domain (SD) and swept-source (SS) optical coherence tomography angiography (OCTA) and measured using an automated detection algorithm. METHODS. Patients diagnosed with CNV were imaged by SD-OCTA and SS-OCTA systems using 3 x 3-mm and 6 x 6-mm scans. The complex optical microangiography (OMAG(C)) algorithm was used to generate the OCTA images. Optical coherence tomography A datasets for imaging CNV were derived by segmenting from the outer retina to 8 mu m below Bruch's membrane. An artifact removal algorithm was used to generate angiograms free of retinal vessel projection artifacts. An automated detection algorithm was developed to quantify the size of the CNV. Automated measurements were compared with manual measurements. Measurements from SD-OCTA and SS-OCTA instruments were compared as well. RESULTS. Twenty-seven eyes from 23 subjects diagnosed with CNV were analyzed. No significant differences were detected between manual and automatic measurements: SD-OCTA 3 x 3-mm (P = 0.61, paired t-test) and 6 x 6-mm (P = 0.09, paired t-test) scans and the SS-OCTA 3 x 3-mm (P = 0.41, paired t-test) and 6 x 6-mm (P = 0.16, paired t-test) scans. Bland-Altman analyses were performed to confirm the agreement between automatic and manual measurements. Mean lesion sizes were significantly larger for the SS-OCTA images compared with the SD-OCTA images: 3 3 3-mm scans (P = 0.011, paired sample t-test) and the 6 x 6-mm scans (P = 0.021, paired t-test). CONCLUSIONS. The automated algorithm measurements of CNV were in agreement with the hand-drawn measurements. On average, automated SS-OCTA measurements were larger than SD-OCTA measurements and consistent with the results from using hand-drawn measurements.
URI: https://repositorio.unifesp.br/handle/11600/55012
ISSN: 0146-0404
Other Identifiers: http://dx.doi.org/10.1167/iovs.16-20977
Appears in Collections:Artigo

Files in This Item:
File SizeFormat 
WOS000398089000019.pdf1.05 MBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.