Please use this identifier to cite or link to this item: https://repositorio.unifesp.br/handle/11600/31119
Title: Vital Dyes and Light Sources for Chromovitrectomy: Comparative Assessment of Osmolarity, pH, and Spectrophotometry
Authors: Costa, Elaine de Paula Fiod [UNIFESP]
Rodrigues, Eduardo Buchele [UNIFESP]
Farah, Michel Eid [UNIFESP]
Dib, Eduardo [UNIFESP]
Penha, Fernando Marcondes [UNIFESP]
Magalhaes, Octaviano [UNIFESP]
Furlani, Bruno Albuquerque [UNIFESP]
Lima Filho, Acácio Alves de Souza [UNIFESP]
Miranda, Antonio de [UNIFESP]
Maia, Mauricio [UNIFESP]
Universidade Federal de São Paulo (UNIFESP)
Issue Date: 1-Jan-2009
Publisher: Assoc Research Vision Ophthalmology Inc
Citation: Investigative Ophthalmology & Visual Science. Rockville: Assoc Research Vision Ophthalmology Inc, v. 50, n. 1, p. 385-391, 2009.
Abstract: PURPOSE. To investigate the in vitro pH, osmolarity, spectral, and photostability properties of nine vital dyes for vitreoretinal surgery.METHODS. Nine dyes-indocyanine green (ICG), trypan blue (TB), brilliant blue (BriB), bromophenol blue (BroB), Congo red (CR), light green (LG), fast green (FG), indigo carmine (IC) and Evans blue (EB)-diluted in three solvents (saline solution, glucose 5%, and water) were tested for osmolarity and pH. Spectrophotometry was used to determine absorbance properties of 27 solutions. Irradiance emission spectra of seven endoillumination light sources and fiber-optics were compared with dye absorbance curves.RESULTS. Dye osmolarity in saline solution and glucose 5% varied widely (257-385 mOsm) and was lower (0-54 mOsm) when dyes were dissolved in water. Dyes diluted in three solvents showed pH values varying from 2.6 to 9.85. ICG, LG, TB, BroB, CR, and IC demonstrated different absorbances, depending on the solvent. BriB and FG showed similar absorbance curves with different solvents. Spectrophometric analysis showed that all dyes except ICG had remarkable spectral overlap with the light sources. Among endoillumination fiber-optics, overlap was greatest with dual-output illumination with an integrated laser pathway and least with a mercury vapor lamp.CONCLUSIONS. Vital dyes showed variable osmolarity and pH, which also depended on the solvent used. Interaction of light from endoillumination source and vital dye may increase or decrease the risk for toxicity, making appropriate selection of both a desirable way to minimize the risk for phototoxic effects. (Invest Ophthalmol Vis Sci. 2009; 50: 385-391) DOI: 10.1167/iovs.08-2285
URI: http://repositorio.unifesp.br/handle/11600/31119
ISSN: 0146-0404
Other Identifiers: http://dx.doi.org/10.1167/iovs.08-2285
Appears in Collections:Em verificação - Geral

Files in This Item:
There are no files associated with this item.


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.